FDA flags overheating power connections on Tablo dialysis equipment
Reports involve both generations of power cord. Outset Medical asked customers to check the connection on its dialysis console and cart before each use.
Original-source briefing · AI-written · Edited by Codex
Source event: 2026-11-18. Published: 2026-11-19T12:51:00.175Z. Updated: 2026-11-19T12:51:00.175Z.
What happened
Primary-source statement · The U.S. Food and Drug Administration issued an early alert on November 18 after Outset Medical received reports that power connections were overheating on its Tablo dialysis equipment. The issue involves two devices: the Tablo Console and the TabloCart with Prefiltration Drawer. In a November 10 letter to customers, Outset recommended continued use with revised checks before each use. It told customers to take a unit out of service and contact technical support if they found signs such as a loose connector, charring or melting. [U.S. Food and Drug Administration]
Why it matters
Primary-source statement · The reports involve both the original cord with clear ends and the newer cord with gray ends. Outset traced the heat to the point where the cord plugs into the equipment. According to the company, a connector that moves in its socket can increase electrical resistance and generate heat. The FDA said contact with an overheated connector could cause a burn. If the console loses power, dialysis could be delayed, interrupted or missed. Outset asked facilities to share the notice with staff, organizations that received transferred units and patients served by home programs. [U.S. Food and Drug Administration]
What remains unknown
Primary-source statement · As of November 10, Outset had reported no serious injuries or deaths associated with the issue, according to the FDA. The agency’s notice, current as of November 19, gives neither the number of overheating reports nor the number of units in use, so it cannot establish how often this occurs. The FDA says its review is continuing. [U.S. Food and Drug Administration]

Read the infographic as text
Two cords, one connection point. The FDA lists the original clear-ended cord, PN-0005746, and current gray-ended cord, PN-0008460. Reports involve the power inlet on the Tablo Console and TabloCart with Prefiltration Drawer. The connection diagram is schematic and is not to scale.
Two possible consequences. Outset says movement at a loose connection increases resistance and generates heat. The FDA identifies a possible contact burn and possible loss of console power. A power interruption could delay, interrupt or prevent a treatment.
Which date goes with which finding. Outset sent its customer letter on November 10. As of that date the company had reported no serious injuries or deaths associated with this issue. The FDA posted its early alert on November 18. The page was current November 19 and gives no count of overheating reports or total units in use.
Graphic sources: U.S. Food and Drug Administration
The developing record.
Earlier entries preserve what was published at that time; later entries may revise them.
2026-11-19T12:51:00.175Z · The FDA issued an early alert about overheating power connections and published Outset Medical’s revised use instructions.
FDA flags overheating power connections on Tablo dialysis equipment
Reports involve both generations of power cord. Outset Medical asked customers to check the connection on its dialysis console and cart before each use.
What happened
The U.S. Food and Drug Administration issued an early alert on November 18 after Outset Medical received reports that power connections were overheating on its Tablo dialysis equipment. The issue involves two devices: the Tablo Console and the TabloCart with Prefiltration Drawer. In a November 10 letter to customers, Outset recommended continued use with revised checks before each use. It told customers to take a unit out of service and contact technical support if they found signs such as a loose connector, charring or melting. [U.S. Food and Drug Administration]
Why it matters
The reports involve both the original cord with clear ends and the newer cord with gray ends. Outset traced the heat to the point where the cord plugs into the equipment. According to the company, a connector that moves in its socket can increase electrical resistance and generate heat. The FDA said contact with an overheated connector could cause a burn. If the console loses power, dialysis could be delayed, interrupted or missed. Outset asked facilities to share the notice with staff, organizations that received transferred units and patients served by home programs. [U.S. Food and Drug Administration]
What remains unknown
As of November 10, Outset had reported no serious injuries or deaths associated with the issue, according to the FDA. The agency’s notice, current as of November 19, gives neither the number of overheating reports nor the number of units in use, so it cannot establish how often this occurs. The FDA says its review is continuing. [U.S. Food and Drug Administration]

Read the infographic as text
Two cords, one connection point. The FDA lists the original clear-ended cord, PN-0005746, and current gray-ended cord, PN-0008460. Reports involve the power inlet on the Tablo Console and TabloCart with Prefiltration Drawer. The connection diagram is schematic and is not to scale.
Two possible consequences. Outset says movement at a loose connection increases resistance and generates heat. The FDA identifies a possible contact burn and possible loss of console power. A power interruption could delay, interrupt or prevent a treatment.
Which date goes with which finding. Outset sent its customer letter on November 10. As of that date the company had reported no serious injuries or deaths associated with this issue. The FDA posted its early alert on November 18. The page was current November 19 and gives no count of overheating reports or total units in use.