React Health warns of inaccurate pressure reports from 7,533 breathing devices

A firmware correction changes what affected Luna G3 devices record going forward. Clinicians reviewing earlier reports can still see inaccurate pressure readings.

Original-source briefing · AI-written · Edited by Codex

Source event: 2026-11-19. Published: 2026-11-19T19:55:55.072Z. Updated: 2026-11-19T19:55:55.072Z.

What happened

Primary-source statement · A clinician reviewing a Luna G3 breathing device from a distance could see a steady pressure reading even though the pressure varied during treatment. React Health described the problem in a November 17 update that the U.S. Food and Drug Administration posted November 19. The company said its correction covers 7,533 devices distributed in the United States and Puerto Rico. Only specific serial numbers of model LG3800-30VT are affected. The reporting error occurs when a feature called Targeted VT is switched on. [U.S. Food and Drug Administration]

Why it matters

Primary-source statement · The device can adjust pressure during treatment to reach a volume set by the clinician. The pressure statistics it saves to a memory card or sends to React Health Connect, the company’s cloud service, can miss those changes. React Health warned that a clinician relying on that history could adjust treatment based on a misleading reading and delay appropriate therapy. The company asks patients using the affected feature to contact their clinician or equipment provider promptly and to continue prescribed therapy until they receive guidance. [U.S. Food and Drug Administration]

What remains unknown

Primary-source statement · React Health said it had received 15 complaints and described two serious adverse reports. Its notice does not describe the two outcomes or say how many patients used the feature. The company first communicated the correction on October 8. It says firmware version G3-2.00.80 corrects future pressure reports. Installing it cannot reconstruct what pressure a device delivered during an earlier session. [U.S. Food and Drug Administration]

A flow diagram follows pressure data from Targeted VT through the memory card and the cloud into a clinician’s report. A timeline marks the October notice, November update and FDA posting. The firmware fixes new records and leaves old readings unchanged.
How a pressure reading can become misleading on its way from an affected Luna G3 to a clinician’s historical report. · ICU Message graphic based on React Health’s November 17 correction update, posted by the FDA November 19, 2026.
Read the infographic as text

One setting, two routes to the report. On affected model LG3800-30VT units with Targeted VT enabled, pressure can vary during therapy. The statistics saved to the device SD card or transmitted to React Health Connect can instead show a steady value. Clinicians may review either source.

Which records change. React Health says firmware G3-2.00.80 corrects recording going forward. Previously recorded pressure data from the affected setting cannot be reconstructed by the update.

Who and when. React Health lists 7,533 distributed devices with specific serial numbers in the U.S. and Puerto Rico. It first communicated the correction October 8, issued its update November 17, and the FDA posted that company update November 19.

Graphic sources: U.S. Food and Drug Administration

The developing record.

Earlier entries preserve what was published at that time; later entries may revise them.

2026-11-19T19:55:55.072Z · React Health explained which Luna G3 pressure records can be inaccurate and how the firmware correction affects future reports.

React Health warns of inaccurate pressure reports from 7,533 breathing devices

A firmware correction changes what affected Luna G3 devices record going forward. Clinicians reviewing earlier reports can still see inaccurate pressure readings.

What happened

A clinician reviewing a Luna G3 breathing device from a distance could see a steady pressure reading even though the pressure varied during treatment. React Health described the problem in a November 17 update that the U.S. Food and Drug Administration posted November 19. The company said its correction covers 7,533 devices distributed in the United States and Puerto Rico. Only specific serial numbers of model LG3800-30VT are affected. The reporting error occurs when a feature called Targeted VT is switched on. [U.S. Food and Drug Administration]

Why it matters

The device can adjust pressure during treatment to reach a volume set by the clinician. The pressure statistics it saves to a memory card or sends to React Health Connect, the company’s cloud service, can miss those changes. React Health warned that a clinician relying on that history could adjust treatment based on a misleading reading and delay appropriate therapy. The company asks patients using the affected feature to contact their clinician or equipment provider promptly and to continue prescribed therapy until they receive guidance. [U.S. Food and Drug Administration]

What remains unknown

React Health said it had received 15 complaints and described two serious adverse reports. Its notice does not describe the two outcomes or say how many patients used the feature. The company first communicated the correction on October 8. It says firmware version G3-2.00.80 corrects future pressure reports. Installing it cannot reconstruct what pressure a device delivered during an earlier session. [U.S. Food and Drug Administration]

A flow diagram follows pressure data from Targeted VT through the memory card and the cloud into a clinician’s report. A timeline marks the October notice, November update and FDA posting. The firmware fixes new records and leaves old readings unchanged.
How a pressure reading can become misleading on its way from an affected Luna G3 to a clinician’s historical report. · ICU Message graphic based on React Health’s November 17 correction update, posted by the FDA November 19, 2026.
Read the infographic as text

One setting, two routes to the report. On affected model LG3800-30VT units with Targeted VT enabled, pressure can vary during therapy. The statistics saved to the device SD card or transmitted to React Health Connect can instead show a steady value. Clinicians may review either source.

Which records change. React Health says firmware G3-2.00.80 corrects recording going forward. Previously recorded pressure data from the affected setting cannot be reconstructed by the update.

Who and when. React Health lists 7,533 distributed devices with specific serial numbers in the U.S. and Puerto Rico. It first communicated the correction October 8, issued its update November 17, and the FDA posted that company update November 19.

Graphic sources: U.S. Food and Drug Administration

Original sources

Medical Device Correction Update Luna® G3 BPAP System (Model LG3800-30VT) · Issuer: React Health (3B Medical, Inc.) · Published 2026-11-19