FDA warns certain breathing-tube connectors could separate during use
The Intersurgical alert identifies specific products and lots for quarantine. The possible harm is interrupted ventilation; the manufacturer’s report of no serious injuries is dated November 3.
Original-source briefing · AI-written · Edited by Codex
Source event: 2026-11-10. Published: 2026-11-16T13:54:59.773Z. Updated: 2026-11-16T13:54:59.773Z.
What happened
Primary-source statement · The FDA is warning about certain Intersurgical 22mm Smoothbore breathing tubes and systems because soft connectors could separate from the tubing. Its November 10 early alert says the manufacturer traced the problem to too little bonding solvent in some products, after disconnections were reported outside the United States. Intersurgical sent affected customers a letter on November 3. The FDA tells facilities to stop using and quarantine the affected lots, return them, and pass the notice to users and customers who received them. [U.S. Food and Drug Administration] [U.S. Food and Drug Administration]
Why it matters
Primary-source statement · These tubes connect patients to machines that help them breathe. A disconnected connector could interrupt or reduce ventilation, the FDA says, potentially leading to low oxygen and serious harm. The alert’s product table lists 11 product references and 38 product-and-lot combinations, counted by ICU Message. Those are inventory identifiers, not numbers of patients or devices. Matching both fields matters: the notice does not say every Intersurgical breathing system is affected. [U.S. Food and Drug Administration]
What remains unknown
Evidence limits · According to the FDA, Intersurgical had reported no serious injuries or deaths associated with this issue as of November 3. That statement does not establish the outcome for every patient or describe reports received since then. The FDA says it is still reviewing the potentially high-risk issue. Its early-alert program communicates such warnings while regulatory review may still be underway; an early alert should not be read as a completed recall classification. The reviewed notice does not give a total number of affected patients or devices. [U.S. Food and Drug Administration] [U.S. Food and Drug Administration]
This report describes a dated FDA device alert and attributed manufacturer findings. It is not an individual medical assessment or instructions for changing a patient’s breathing support.

Read the infographic as text
Possible connector failure. Intersurgical reported too little bonding solvent in some products. If a soft connector separates from the tube, ventilation may be interrupted or reduced, potentially lowering oxygen and causing serious harm. The drawing is schematic and not to scale.
Exact product-and-lot inventory. 5018000: 32511811. 5025000: 32522651. 5025310: 32526115. 5026000: 32525551. 5183000: 32527207. 5191000: 32520670, 32522190, 32522802. 5191850: 32521551, 32521655. 5194850: 32518127. 5804000: 32508034, 32509327, 32510909, 32516895, 32518659, 32519175, 32525543. 5805000: 32516315, 32517877. 5805030: 32505469, 32506841, 32513929, 32514276, 32515579, 32518484, 32519342, 32519468, 32519979, 32522143, 32522415, 32523651, 32525498, 32525628, 32526171, 32526270, 32527299, 32528409. These 11 product references contain 38 listed product/lot combinations. They are not counts of individual devices or patients and do not express relative risk. Check the linked FDA table for the complete product descriptions.
Dates and uncertainty. November 3, 2026: manufacturer customer letter and report of no serious injuries or deaths associated with the issue as of that date. November 10: FDA early alert. The FDA says review continues; a dated absence of reports does not establish zero risk.
FDA inventory instructions. Identify exact affected products and lots, stop using and quarantine them, return affected products to Intersurgical, and pass the notice to users and downstream recipients. This report does not provide patient-specific treatment instructions.
Graphic sources: U.S. Food and Drug Administration · U.S. Food and Drug Administration · U.S. Food and Drug Administration
The developing record.
Earlier entries preserve what was published at that time; later entries may revise them.
2026-11-16T13:54:59.773Z · Initial report on the November 10 FDA Intersurgical early alert, with product-lot inventory, connector schematic and dated evidence timeline.
FDA warns certain breathing-tube connectors could separate during use
The Intersurgical alert identifies specific products and lots for quarantine. The possible harm is interrupted ventilation; the manufacturer’s report of no serious injuries is dated November 3.
What happened
The FDA is warning about certain Intersurgical 22mm Smoothbore breathing tubes and systems because soft connectors could separate from the tubing. Its November 10 early alert says the manufacturer traced the problem to too little bonding solvent in some products, after disconnections were reported outside the United States. Intersurgical sent affected customers a letter on November 3. The FDA tells facilities to stop using and quarantine the affected lots, return them, and pass the notice to users and customers who received them. [U.S. Food and Drug Administration] [U.S. Food and Drug Administration]
Why it matters
These tubes connect patients to machines that help them breathe. A disconnected connector could interrupt or reduce ventilation, the FDA says, potentially leading to low oxygen and serious harm. The alert’s product table lists 11 product references and 38 product-and-lot combinations, counted by ICU Message. Those are inventory identifiers, not numbers of patients or devices. Matching both fields matters: the notice does not say every Intersurgical breathing system is affected. [U.S. Food and Drug Administration]
What remains unknown
According to the FDA, Intersurgical had reported no serious injuries or deaths associated with this issue as of November 3. That statement does not establish the outcome for every patient or describe reports received since then. The FDA says it is still reviewing the potentially high-risk issue. Its early-alert program communicates such warnings while regulatory review may still be underway; an early alert should not be read as a completed recall classification. The reviewed notice does not give a total number of affected patients or devices. [U.S. Food and Drug Administration] [U.S. Food and Drug Administration]

Read the infographic as text
Possible connector failure. Intersurgical reported too little bonding solvent in some products. If a soft connector separates from the tube, ventilation may be interrupted or reduced, potentially lowering oxygen and causing serious harm. The drawing is schematic and not to scale.
Exact product-and-lot inventory. 5018000: 32511811. 5025000: 32522651. 5025310: 32526115. 5026000: 32525551. 5183000: 32527207. 5191000: 32520670, 32522190, 32522802. 5191850: 32521551, 32521655. 5194850: 32518127. 5804000: 32508034, 32509327, 32510909, 32516895, 32518659, 32519175, 32525543. 5805000: 32516315, 32517877. 5805030: 32505469, 32506841, 32513929, 32514276, 32515579, 32518484, 32519342, 32519468, 32519979, 32522143, 32522415, 32523651, 32525498, 32525628, 32526171, 32526270, 32527299, 32528409. These 11 product references contain 38 listed product/lot combinations. They are not counts of individual devices or patients and do not express relative risk. Check the linked FDA table for the complete product descriptions.
Dates and uncertainty. November 3, 2026: manufacturer customer letter and report of no serious injuries or deaths associated with the issue as of that date. November 10: FDA early alert. The FDA says review continues; a dated absence of reports does not establish zero risk.
FDA inventory instructions. Identify exact affected products and lots, stop using and quarantine them, return affected products to Intersurgical, and pass the notice to users and downstream recipients. This report does not provide patient-specific treatment instructions.
Graphic sources: U.S. Food and Drug Administration · U.S. Food and Drug Administration · U.S. Food and Drug Administration