B. Braun changes instructions after reports of IV valves staying open
An FDA early alert describes leakage after disconnection of certain Caresite Micro valves. Its linked product list identifies 79 batches across eight material numbers.
Original-source briefing · AI-written · Edited by Codex
Source event: 2026-11-19. Published: 2026-11-22T15:04:00.527Z. Updated: 2026-11-22T15:04:00.527Z.
What happened
Primary-source statement · B. Braun Medical told health care facilities to check certain Caresite Micro needle-free valves after a connector is disconnected. Customers had reported that the piston inside the valve sometimes stayed depressed, allowing fluid to leak or flow back. The company sent its correction letter on November 13. In a November 19 early alert, the FDA said the instructions call for replacing an affected device if staff see leakage, backflow or a valve that stays open. [FDA]
Why it matters
Primary-source statement · The device gives clinicians a needle-free way to administer blood, IV fluids or medicines. The FDA says a valve that stays open can allow air to enter or interrupt a prescribed infusion. Its alert identifies the valve component as S5404000. For inventory checks, the linked spreadsheet uses different identifiers. It pairs a material number with a batch number on each row. ICU Message counted 79 distinct batches across eight material numbers. Material 354267 appears on 35 rows and 354265 on 27. The other six account for the remaining 17. [FDA] [FDA]
What remains unknown
Evidence limits · B. Braun reported finding missing silicone lubricant at the piston and identified insufficient lubrication as a possible reason for sticking. The alert gives no rate of valve failure and no count of affected patients. It says the company had reported no serious injuries or deaths associated with the issue as of November 3, ten days before the customer letter. The FDA says its review is ongoing. [FDA] [FDA]
This report describes a manufacturer correction addressed to health care facilities. The workbook counts batches. It does not establish how many valves failed or how many patients were exposed.

Read the infographic as text
Component and material numbers. S5404000 identifies the Caresite Micro valve component. The workbook identifies eight material numbers for the affected products. Material numbers, batch numbers and patient counts measure different things.
How the 79 batches are distributed. Material 354267 has 35 batches, 354265 has 27, 354268 has nine, 354266 has three and 354260 has two. Materials 354261, 354262 and 354263 have one each. All bars start at zero. A batch can contain more than one device.
The three dates in the alert. November 3 was the cutoff for the manufacturer’s report of no serious injuries or deaths. B. Braun sent its customer letter on November 13. FDA posted the early alert on November 19. The workbook gives no failure rate.
The developing record.
Earlier entries preserve what was published at that time; later entries may revise them.
2026-11-22T15:04:00.527Z · FDA reported revised instructions for Caresite Micro valves; its linked workbook lists 79 affected batches.
B. Braun changes instructions after reports of IV valves staying open
An FDA early alert describes leakage after disconnection of certain Caresite Micro valves. Its linked product list identifies 79 batches across eight material numbers.
What happened
B. Braun Medical told health care facilities to check certain Caresite Micro needle-free valves after a connector is disconnected. Customers had reported that the piston inside the valve sometimes stayed depressed, allowing fluid to leak or flow back. The company sent its correction letter on November 13. In a November 19 early alert, the FDA said the instructions call for replacing an affected device if staff see leakage, backflow or a valve that stays open. [FDA]
Why it matters
The device gives clinicians a needle-free way to administer blood, IV fluids or medicines. The FDA says a valve that stays open can allow air to enter or interrupt a prescribed infusion. Its alert identifies the valve component as S5404000. For inventory checks, the linked spreadsheet uses different identifiers. It pairs a material number with a batch number on each row. ICU Message counted 79 distinct batches across eight material numbers. Material 354267 appears on 35 rows and 354265 on 27. The other six account for the remaining 17. [FDA] [FDA]
What remains unknown
B. Braun reported finding missing silicone lubricant at the piston and identified insufficient lubrication as a possible reason for sticking. The alert gives no rate of valve failure and no count of affected patients. It says the company had reported no serious injuries or deaths associated with the issue as of November 3, ten days before the customer letter. The FDA says its review is ongoing. [FDA] [FDA]

Read the infographic as text
Component and material numbers. S5404000 identifies the Caresite Micro valve component. The workbook identifies eight material numbers for the affected products. Material numbers, batch numbers and patient counts measure different things.
How the 79 batches are distributed. Material 354267 has 35 batches, 354265 has 27, 354268 has nine, 354266 has three and 354260 has two. Materials 354261, 354262 and 354263 have one each. All bars start at zero. A batch can contain more than one device.
The three dates in the alert. November 3 was the cutoff for the manufacturer’s report of no serious injuries or deaths. B. Braun sent its customer letter on November 13. FDA posted the early alert on November 19. The workbook gives no failure rate.
Original sources
FDA early alert on B. Braun Caresite Micro Luer Access Devices · Issuer: U.S. Food and Drug Administration · Updated 2026-11-19
Full list of affected Caresite Micro products and batches · Issuer: U.S. Food and Drug Administration · Retrieved (undated original) 2026-11-22